Print Back
  • Accepts a variety of standard process signals and precisely scales them into any desired unit of measurement
  • Single input or dual input
  • Employs advanced technology for stable, drift-free readout
  • Field upgradeable with plug-in option cards; including 4 set point alarms, analog output and serial communication
  • 16 point scaling for non-linear processes
  • Max. and min. value display
Unit of Measure

Specifications

Type

Digital Indicator

Series

Single Input

Process Display

5 digit, 0.56" Red LED

Operating Temperature [Min]

0 ºC32 ºF

Operating Temperature [Max]

50 ºC122 ºF
Accuracy1 <0.03% of full scale, 64 ºF to 82 ºF (18 ºC to 50 ºC)
<0.12% of full scale, 32 ºF to 122 ºF (0 ºC to 50 ºC)

Storage Temperature

-40 to 140 ºF-40 to 60 ºC

Effect

≤0.1 %

Full Scale Temperature

18 ºF10 ºC

Environmental Rating

National Electrical Manufacturers Association (NEMA) 4, National Electrical Manufacturers Association (NEMA) 4X, Ingress Protection (IP65) (Front of Display Only)

Input Signals Current

4 to 20 mA

Alternating Current (AC) Voltage

115 V230 V

Direct Current (DC) Transmitter Power Supply Voltage

24 V

Key Pad

3 programmable function keys, 5 total

Power Requirement

AC version 85 Vac to 250 Vac, 50/60 Hz, 15 VA, DC version 11 Vdc to 36 Vdc, 11 W

Electrical Connection

Screw terminals at back of case

Memory

Electrically Erasable Programmable Read-Only Memory (EEPROM)

Non-Condensing Humidity

85 %

Vibration

5 to 150 Hz2 g

Shock

25 g

Normal Rejection Mode

60 dB

Normal Rejection Mode Frequency

50 Hz60 Hz

Altitude

Up to 2,000 m

Update Rate

200 ms

Selectable Filtering

0 to 25 s

Linearization

2 to 16 points, selectable

Reading [Min/Max]

Captures and stores min/max measurement

User Inputs

3 programmable, 30 Vdc max. continuous

Sensor Excitation

Built-in 24 Vdc

Totalizer Time Base

Second, Minute, Hour or Day

Totalizer Time Batch

Can accumulate from user input

Electromagnetic Rating

European Conformity (CE) Compliant to Electromagnetic Compatibility (EMC) Norm European Standards (EN) 61326-1, Radio Frequency Interference (RFI), Electromagnetic Interference (EMI) and Electrostatic Discharge (ESD) Protection

Weight

10.4 oz

Warranty

2 y

Options

Resistance Temperature Device (RTD), Thermocouple

Option Cards

Analog Output Option Card

Limitations Which Apply Are

Limitations Which Apply Are

These units must be operated within the catalogued environmental and application parameters. Determination of failure will be made by NOSHOK, Inc.'s equipment and personnel or a certified test facility specializing in this type of evaluation.

In keeping with and for purposes of product and/or manufacturing process improvements, NOSHOK, Inc. reserves the right to make design changes without prior notice.

Application

Application

  • Process indication
  • On/Off pump controls
  • Compressor controls
  • Safety or shutdown alarms
  • Local indication with signal
  • Local indication with signal retransmission to computer control system

Additional Information

Additional Information

  • Easy menu-driven programming
  • NEMA 4X/IP65 sealed front bezel
  • Optional PC software available for configuration
  • AC or DC input power
  • Signal totalizer for batch weighing or other timed input processes
  • Programmable signal response time
  • Standard DIN panel cutout
  • CE compliant to suppress RFI, EMI, and ESD
  • Serial communication (RS232 or RS485)
  • DeviceNet communication
  • MODBUS communication (RS485 RTU and ASCII MODBUS)
  • Profibus communication
  • Analog output (0-20 mA, 4-20 mA or 0-10 Vdc)
  • Dual or Quad Relay (NPN or PNP type)
  • USB programming

Options and Accessories

2000/2100 Series Smart System Intelligent Digital Indicator Option Cards

  • Analog Output Option Card
  • Quad Relay Option Card
  • Quad NPN-OC Option Card
  • Quad PNP-OC Option Card
  • RS232-C Serial Communications Option Card
  • RS 485 Serial Communications Option Card

AC to DC Retrofit Package

  • Easily integrates with non-metallic seal and transmitter/switch assemblies to operate with 110/240 VAC systems
  • Solid state relay - 10A at 250 VAC
  • DIN rail mountable with NEMA 4X enclosure

NEMA 4 Enclosure - 1950 Series Digital Indicator

  • NEMA 4X/IP65 enclosures (sealed bezel only) standard
  • Black painted steel enclosures (ENC1) are available
  • 1 Max. error on complete span. Error at calibration point ≤0.1 ºC